Abstract
Some potential applications of the isotopes36Cl and236U require higher sensitivity than is normally required or achieved. For36Cl, one such application is the measurement of the36Cl/Cl ratio of seawater. We show that, with appropriate precautions to reduce the effect of ion-source memory, it is possible to reach blank levels below 10-16in the36Cl/Cl ratio. Measurements of a series of seawater and sea salt samples yield a blank-corrected36Cl/Cl ratio of present-day sea water of 0.71 ± 0.08 × 10-15. For236U, one potential application is as a prospecting tool, where sensitivities in the236U/238U ratio at the 10-13level are desirable. We demonstrate that this sensitivity can be achieved for 24 MeV uranium ions with time-of-flight detection over a 6.0 m flight path.