Measurement of pH at elevated temperatures using the optical indicator acridine

Loading...
Thumbnail Image

Date

Journal Title

Journal ISSN

Volume Title

Publisher

Abstract

A new method of measuring the pH of high-temperature aqueous solutions using optical indicators was developed. The absorbance spectrum of acridine changes as a function of pH, allowing its use as an optical indicator for pH. The pKa of acridine were experimentally determined in KOH-AcOH pH-buffer solutions at temperatures from 5 to 250°C at vaporsaturated pressures: {Mathematical expression} where T is in Kelvin. The technique is demonstrated by experimentally determining the pKa of benzoic acid at temperatures up to 250°C where the result was {Mathematical expression}