36Cl accelerator mass spectrometry with a bespoke instrument
Loading...
Date
Journal Title
Journal ISSN
Volume Title
Publisher
Abstract
Cosmogenic 36Cl analysis by accelerator mass spectrometry (AMS) is a valuable environmental and geological sciences research tool. Overcoming the stable nuclide 36S isobar interfering with measurement is challenging, however. Traditionally this has required large accelerators, but following recent technical advances it is now possible with ∼30 MeV ion energies. Consequently 5 MV or even smaller modern bespoke spectrometers are now 36Cl-capable, increasing accessibility and promoting wider and more varied 36Cl use. However, the technical ability to identify 36Cl ions is quite distinct from demonstrated high-performance AMS. Such is the theme of this paper. We present a systematic analysis of the accurate measurement of sample radioisotope relative to the stable chlorine, the normalisation of the measured ratio and correction for remaining 36S interference, all combined with the use of stable-isotope dilution to determine sample Cl concentration to begin with. We conclude by showing that repeated analyses support our claims for routine 3% 36Cl-AMS data. Accordingly, the modest SUERC spectrometer well competes with the performance of larger longer-established instruments, and the results may be quite generic for modern bespoke instruments.