The use of a one-dimensional position-sensitive detector for diffuse X-ray scattering measurements is described. Calibration procedures for scattering angle and intensity measurements are discussed. Some nonuniformities have been found in the counting efficiency as a function of distance along the detector. A procedure is described for measuring the diffuse scattering in a section of reciprocal space.