Measurement of sulfur isotope ratios in micrometer-sized samples by NanoSIMS

Loading...
Thumbnail Image

Journal Title

Journal ISSN

Volume Title

Publisher

Abstract

Sulfur isotope ratios of atmospheric aerosol particles can provide detailed information with regard to the origin and the transport of sulfur in the environment. The new Cameca NanoSIMS 50 ion microprobe technique permits analysis of individual aerosol particles with volumes down to 0.5 μm 3 and a precision for δ 34 S of 3-10‰ (2σ). This technique will set new standards in the analysis of isotope ratios in atmospheric aerosol. For the first time it is possible to directly compare chemical and isotopic composition of individual aerosol particles, identify internal and external mixtures and investigate reactions of anthropogenic gases with natural aerosol such as sea salt and mineral dust.