Small changes to our ion source and gas ionization detector have significantly improved Cl measurement by reducing source memory and increasing interference rejection. Gas stripped low energy 30 MeV 36Cl5+ ions are still efficiently transported to the detector but 36S vs. 36Cl separation is improved by an order of magnitude. Accordingly 36Cl/Cl background is <10-15 before additional interference correction that is also newly automated. 35Cl- currents have increased to 30 μA and 36Cl/Cl inter-cathode repeatability is 3%.