A precision apparatus, with solid phase micro-extraction monitoring capability, for incorporation studies of gaseous precursors into insect-derived metabolites
Loading...
Date
Journal Title
Journal ISSN
Volume Title
Publisher
Abstract
An apparatus is described that facilitates the determination of incorporation levels of isotope-labelled, gaseous precursors into volatile insect-derived metabolites. Atmospheres of varying gas compositions can be generated by evacuation of a working chamber followed by admission of the required levels of component gases, using a precision, digitised pressure read-out system. Insects such as fruit-flies are located initially in a small introduction chamber, from which migration can occur downwards into the working chamber. The level of incorporation of labelled precursors is continuously assayed by the Solid Phase Micro Extraction (SPME) technique and GC-MS analyses. Experiments with both Bactrocera species (fruit-flies) and a parasitoid wasp, Megarhyssa nortoni nortoni (Cresson) and 18O-labelled dioxygen illustrate the utility of this system. The isotope effects of 18O on the carbon-13 NMR spectra of 1,7-dioxaspiro[5,5]undecane are also described.
Description
Citation
Collections
Source
Arkivoc