Abstract
The joint ANSTO/ANU 36Cl program has now measured more than 700 samples from many different locations. During the course of this work, a variety of contamination problems have affected a small number of results which have provided valuable information on the effects of ion source cross-talk, sample preparation and storage procedures and sources of high-36Cl material. A sample of Weeks Island halite is processed along with every batch of field samples and the observed ratio provides a clear distinction between normal batches and those subject to contamination. Over three years, the long-term average ratio of 36Cl Cl for normal halite samples is (1 ± 1) × 10-15. The sample handling procedures developed during the course of this work provide a useful guide to the techniques that must be used to achieve the sensitivity limits which are potentially available using AMS.
Description
Keywords
Citation
Collections
Source
Nuclear Instruments and Methods in Physics Research, Section B: Beam Interactions with Materials and Atoms