Low volume microwave digestion for the determination of selenium in marine biological tissues by graphite furnace atomic absorption spectroscopy

Loading...
Thumbnail Image

Journal Title

Journal ISSN

Volume Title

Publisher

Abstract

A microwave digestion method for the determination of marine biological tissues has been developed to allow determination of selenium in small sample sizes (< 0.1 g). The benefits of this technique include maintaining concentrates in extracts without the subsequent over dilution encountered when using larger vessels, increased sample throughput and reduced loss of volatile material. Freeze dried biological material < 0.1 g) and nitric acid (1 ml) were placed into 7 ml screw top Teflon vessels which are completely sealed on capping. Two 7 ml vials were placed into larger 120 ml vessels fitted with a Teflon spacer and 10 ml of distilled water. The effects of microwave power and time, and sample mass on selenium recovery from three marine standard reference materials (NIST SRM 1566a Oyster Tissue, NRCC DORM-1 Dogfish Muscle and NRCC TORT-1 Lobster Hepatopancreas) were examined. The optimum conditions: 600 W, 2 min; 0 W, 2 min; 450 W, 45 min, allowed quantitative recoveries of selenium from these and three other standard reference materials (NRCC DOLT-1 Dogfish liver, NIST RM-50 Albacore tuna and IAEA MA-A-2 fish flesh). Studies on sample mass showed that the analysis of sample masses from 0.025 to 0.1 g gave selenium concentrations within the certified range. Six species of selenium: selenite, selenate, selenomethionine, selenocysteine, selenocystamine, and trimethyl selenonium were added to oyster, dogfish, and lobster tissues. Recoveries were near quantitative for all species (94-105%) except trimethyl selenonium (90-101%).

Description

Citation

Source

Analytica Chimica Acta

Book Title

Entity type

Access Statement

License Rights

Restricted until