Verification of out-of-control situations detected by “average of normal” approach

Loading...
Thumbnail Image

Journal Title

Journal ISSN

Volume Title

Publisher

Abstract

Objectives “Average of normal” (AoN) or “moving average” is increasingly used as an adjunct quality control tool in laboratory practice. Little guidance exists on how to verify if an out-of-control situation in the AoN chart is due to a shift in analytical performance, or underlying patient characteristics. Design and methods Through simulation based on clinical data, we examined 1) the location of the last apparently stable period in the AoN control chart after an analytical shift, and 2) an approach to verify if the observed shift is related to an analytical shift by repeat testing of archived patient samples from the stable period for 21 common analytes. Results The number of blocks of results to look back for the stable period increased with the duration of the analytical shift, and was larger when smaller AoN block sizes were used. To verify an analytical shift, 3 archived samples from the analytically stable period should be retested. In particular, the process is deemed to have shifted if a difference of > 2 analytical standard deviations (i.e. 1:2 s rejection rule) between the original and retested results are observed in any of the 3 samples produced. The probability of Type-1 error (i.e., false rejection) and power (i.e., detecting true analytical shift) of this rule are < 0.1 and > 0.9, respectively. Conclusions The use of appropriately archived patient samples to verify an apparent analytical shift is preferred to quality control materials. Nonetheless, the above findings may also apply to quality control materials, barring matrix effects.

Description

Citation

Source

Clinical Biochemistry

Book Title

Entity type

Access Statement

License Rights

Restricted until